The Application Value of Color Doppler Ultrasound Combined with Ultrasound Elastography in Differentiating the Benign and Malignant Breast Tumors

XUE Hongfang1, CHEN Juan2   

  1. 1. Department of Ultrasound, Xingtai Hospital of Traditional Chinese Medicine, Xingtai 054000, Hebei, P. R. China;
    2. Department of Oncology, Xingtai Hospital of Traditional Chinese Medicine, Xingtai 054000, Hebei, P. R. China
  • Received:2019-10-30 Online:2020-05-15 Published:2020-05-15

摘要: 本文对彩色多普勒超声(CDUs)结合弹性成像技术(UE)在鉴别乳腺良恶性肿瘤中的应用价值进行了分析。选取2018年1月~2019年2月本院乳腺肿瘤患者103例为研究对象,所有患者均给予CDUs、UE检查,以病理检查为对照,分析CDUs、UE及二者结合对乳腺良恶性肿瘤的鉴别价值。病理检查显示,103例乳腺肿瘤患者中,恶性48例(46.60%)、良性55例(53.40%);乳腺肿瘤恶性者CDUs、UE得分明显高于良性者,差异有统计学意义(P<0.05);在鉴别乳腺良恶性肿瘤的敏感度、特异度、准确度中,CDUs为79.17%、72.73%、75.73%,UE为75.00%、65.45%、69.90%,CDUs与UE结合为95.83%、90.91%、93.20%,CDUs结合UE明显高于CDUs、UE,差异有统计学意义(P<0.05)。本文证实CDUs、UE对乳腺良恶性肿瘤具有良好的鉴别价值,且CDUs结合UE的鉴别价值更高。

关键词: 彩色多普勒超声, 超声弹性成像技术, 鉴别, 乳腺良恶性肿瘤

Abstract: The application value of color Doppler ultrasound (CDUs) combined with ultrasound elastography (UE) in differentiating the benign and malignant breast tumors was analyzed in this study. 103 patients with breast tumor in our hospital from January 2018 to February 2019 were selected as the subjects of this study, all patients were examined with CDUs and UE, and the differential value of CDUs, UE and their combination in benign and malignant breast tumors was analyzed. The results showed that, pathological examination showed that, among 103 cases of breast cancer, 48 cases were malignant (46.60%) and 55 cases were benign (53.40%). The CDUs and UE scores in malignant breast tumor patients were significantly higher than those in benign breast tumor patients, the difference was statistically significant (P<0.05). In the sensitivity, specificity and accuracy in differentiating benign and malignant breast tumors, the CDUs were 79.17%, 72.73%, 75.73%, the UE was 75.00%, 65.45%, 69.90%, the CDUs combined with UE was 95.83%, 90.91%, 93.20%, the CDUs combined with UE was significantly higher than the single of the CDUs and UE, the difference was statistically significant (P<0.05). This study confirms that, CDUs and UE have the good value in differentiating benign and malignant breast tumors, and the CDUs combined with UE have higher the differential value.

Key words: color Doppler ultrasound(CDUs), ultrasound elastography(UE), differentiate, benign and malignant breast tumors